Universal AC input / Full range
Built-in active PFC function, PF>0.94
High efficiency up to 89%
Withstand 300VAC surge input for 5 seconds
Protections: Short circuit / Overload / Over voltage / Over temperature
Medical safety approved (MOOP level)
Built-in cooling fan ON-OFF control
Built-in DC OK signal
Built-in remote ON-OFF control
Standby 5V@0.3A
Built-in remote sense function
No load power consumption<0.8W (Note.7)
Current sharing up to 2400W (3+1) (24V,36V,48V)
5 years warranty
SPECIFICATION
MSP-600-3.3 MSP-600-5 MSP-600-7.5 MSP-600-12 MSP-600-15 MSP-600-24 MSP-600-36 MSP-600-48
MODEL
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER
RIPPLE & NOISE (max.) Note.2
OUTPUT VOLTAGE ADJ. RANGE
VOLTAGE TOLERANCE Note.3
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE Note.5
FREQUENCY RANGE
POWER FACTOR (Typ.)
EFFICIENCY (Typ.)
INPUT
ENVIRONMENT
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT
OVER TEMPERATURE
SAFETY STANDARDS
SAFETY &
FUNCTION
EMC
(Note 4)
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
WORKING TEMP.
DC OK SIGNAL
REMOTE CONTROL
WORKING HUMIDITY
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSIONOTHERS
NOTE
PACKING
OVERLOAD
OVER VOLTAGE
PROTECTION
AC CURRENT (Typ.)
3.3V 5V 7.5V 12V 15V 24V 36V 48V
120A 120A 80A 53A 43A 27A 17.5A 13A
0 ~ 120A 0 ~ 120A 0 ~ 80A 0 ~ 53A 0 ~ 43A 0 ~ 27A 0 ~ 17.5A 0 ~ 13A
396W 600W 600W 636W 645W 648W 630W 624W
100mVp-p 100mVp-p 100mVp-p 120mVp-p 150mVp-p 150mVp-p 200mVp-p 240mVp-p
2.8 ~ 3.8V 4.3 ~ 5.8V 6.8 ~ 9V 10.2 ~ 13.8V 13.5 ~ 18V 21.6 ~ 28.8V 28.8 ~ 39.6V 40.8 ~ 55.2V
±2.0% ±2.0% ±2.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0%
±0.5% ±0.5% ±0.5% ±0.3% ±0.3% ±0.2% ±0.2% ±0.2%
±1.0% ±1.0% ±1.0% ±0.5% ±0.5% ±0.5% ±0.5% ±0.5%
1000ms, 50ms/230VAC 2500ms, 50ms/115VAC at full load
16ms/230VAC 16ms/115VAC at full load
85 ~ 264VAC 120 ~ 370VDC
47 ~ 63Hz
PF>0.94/230VAC PF>0.99/115VAC at full load
78.5% 82%
6 ~ 7V 9.4 ~ 10.9V 14.4 ~ 16.8V 18.8 ~ 21.8V 30 ~ 34.8V 41.4 ~ 48.6V 57.6 ~ 67.2V
86% 88% 88% 88% 89% 89%
8.5A/115VAC 5A/230VAC
35A/115VAC 80A/230VAC
105 ~ 135% rated output power
3.96 ~ 4.62V
Protection type : Constant current limiting, recovers automatically after fault condition is removed
Protection type : Shut down o/p voltage, re-power on to recover
Shut down o/p voltage, recovers automatically after temperature goes down
ANSI/AAMI ES60601-1, IEC60601-1 EAC TP TC 004 approved
,
Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3, EAC TP TC 020
Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, EAC TP TC 020
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH℃
-40 ~ +70 (Refer to "Derating Curve")℃
RC+ / RC-: 4 ~ 10V or open = power on ; 0 ~ 0.8V or short = power off
PSU turn on : 3.3 ~ 5.6V ; PSU turn off : 0 ~ 1V
20 ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH non-condensing℃
±℃℃0.03%/ (0 ~ 50 )
10 ~ 500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes
138.7K hrs min. MIL-HDBK-217F (25 )℃
218*105*63.5mm (L*W*H)
File Name:MSP-600-SPEC 01-152018-
■ Features :
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.℃
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on
a mm* mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives.360 360 For guidance on how to
perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com)
5. Derating may be needed under low input voltages. Please check the derating curve for more details.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7. No load power consumption<0.8W when RC+ & RC- (CN100 pin3,4) 0 ~ 0.8V or short.
8. When the input voltage is less than 40VAC, the SPS may exhibit degradation of performance. The final product manufacturers must re-confirm this
deviation that does not affect basic safety or essential performance.
9. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher than 2000m(6500ft).℃℃
MSP-600 series
600W Single Output Medical Type
1.57Kg;8pcs/13.6Kg/1.34CUFT
5V STANDBY
5VSB : 5V@0.3A ; tolerance 5%, ripple : 50mVp-p(max.)±
FAN CONTROL (Typ.)
Load 35 15% or RTH2 50 Fan on±≧℃
EMC IMMUNITY
EMC EMISSION
Earth leakage current < 300 A/264VAC , Touch leakage current < 100 A/264VAC
ISOLATION LEVEL
Primary-Secondary: 2×MOOP, Primary-Earth: 1×MOOP, Secondary-Earth: 1×MOOP