1
CoolPac Slots Power
Medical Approval
IEC60601-1 3rd edi-
tion & IEC60601-1-2
4th edition (EMC)
Industrial Approval
IEC60950 2nd edition
CX06S
CX06M
4
4
600W
600W
-
Yes
Yes
-
Model Vnom (V)
Set Point
Adjust Range
(V)
Imax (A) Power (W)
CmA
CmB
CmC
CmD
5
12
24
48
2.5-6.0
6.0-15.0
15.0-28.0
28.0-58.0
21
15
8.33
4.17
105
180
200
200
NFF Series; No Fan Featured
• 100% natural convection cooled
• No internal or external fans needed
• No base-plate needed
No Acoustic Noise or Vibrations &
Breakthrough Level Reliability
- No fans required – 600W with natural
convection
- MTBF >400000 hours - 25% better,
than todays leading solutions
High Input Surge Protection
- 4KV Line to PE for harsh environments
Reverse Energy Protection
- No blocking diodes required
24W Standby Power
Safety Approved to 5000m Altitude
>93% Efciency
5 Year Warranty
Flexibility
Analog and Digital Management
- PMBus monitoring and control capability
Field Congurable
- Plug and Play Power
Series and Parallel Outputs
- Higher voltages/currents
Mounting Options
- Base/Side and DIN-Rail mounting
Certications
Medical
- IEC60601-1 3rd Edition compliant
- IEC60601-1-2 4th edition EMC compliant
- 2 MOPP
- Dual Fused
- BF rated
- ISO13485
Industrial
- IEC60950, IEC62368-1
- ISO9000
- SEMI F47
Defence/Aero
- MIL810G certied
- MIL461G, CE101, CE102
TYPICAL APPLICATIONS
Medical; Clinical diagnostic equipment,
Medical lasers, Dialysis equipment,
Radiological Imaging, Clinical Chemistry
Industrial; Test and Measurement,
Industrial Machines, Automation equipment,
Printing, Telecommunications,
Audio equipment
Hi Rel; Harsh Industrial Electronics, Radar
(Naval, Ground Based), Communications
Test & Measurement
Cool
CoolX600 NFF Series
600
CoolX CoolPac
CoolX CoolMod
NFF Series
www.excelsys.com
TM
The World’s Only Fanless modular power supply, the CoolX600 NFF Series is
the revolutionary new Convection-Cooled modular power supply from Excelsys. It
provides an incredible 600W in a compact 4.5 x 8.5 x 1U package with no fan.
The silent CoolX600 generates no acoustic noise and offers system designers
best in class performance in efciency (>93%) and unrivalled reliability in
addition to the most comprehensive feature set and specications.
The series comprises two base models. The CX06S is certied to IEC60950 2nd
edition for industrial applications whereas the CX06M carries IEC60601-1 3rd
edition & IEC60601-1-2 4th edition (EMC) for medical applications. The CoolX600
can be populated with up to 4 CoolMods, providing up to 4 isolated DC outputs
ranging from 2.5V to 58.0V. Continuing the Excelsys tradition of exibility, the
CoolX600 is completely user and eld congurable. Outputs can be adjusted
to the required set point voltages and can be congured in parallel or series for
higher current and/or higher voltages. CoolPacs can be paralleled for higher
power and N+1 Redundancy applications.
Stand-out features for medical applications include BF rating (Body Floating),
input dual fusing, 2 x MOPP isolation and <300uA leakage current. Other features
include 4KV input surge immunity, SEMI F47 compliance, MIL810G compliance
and the ability to withstand input voltages of up to 300VAC making it ideal for use
in remote locations and those subject to input voltage disturbances. No fans make
it ideal for acoustic and vibration sensitive applications.
With Digital Communications available, the CoolX600 provides the most
exible, highest specication modular power supply in the market, all backed
up by the Excelsys 5 Year Warranty ensuring quality and the lowest total cost of
ownership
Patented Resonant Technology
2
CoolX600 NFF Series
INPUT
Parameter Conditions/Description Min Nom Max Units
AC Operating Input Range 85 264 VAC
Nominal Input Voltage Range Universal Input 47-440Hz 100 240 VAC
Extended AC Operating Range Maximum for 5 seconds 300 VAC
DC Input voltage range 120 300 VDC
Input Current 90VAC @ 420W 6.0 A
Inrush Current 230VAC @ 600W 25 A
Power Factor 120VAC@ 500W 0.98
Undervoltage Lockout Shutdown 65 74 VAC
Input Fuses Rating Dual Fused (Line and Neutral) 250VAC 8 A
Efciency 230VAC, 600W with 3 x CmC CoolMods 93 %
COOLMOD OUTPUT
Parameter Conditions/Description Min Nom Max Unit
Power Rating CX06: See derating curves 600 W
Minimum Load 0 A
Line Regulation For ±10% change from nominal line ±0.1 %
Load & Cross Regulation For 25% to 75% load change ±0.2 %
Transient Response For 25% to 75% load change 0.5A/uS: Voltage Deviation
Settling Time
4
500
%
µS
Ripple and Noise 100mV or 1.0% pk-pk. 20MHz BW 1 %
Overvoltage Protection Tracking OVP Level
Latching OVP Level
105
125
125
160
%
%
Remote Sense Max. line drop compensation 0.5 VDC
Overshoot 1 %
Rise Time Monotonic 10 ms
Turn-on Delay From AC in
From Global Enable
From CoolMod Enable
800
10
10
ms
ms
ms
Hold-up Time For nominal output voltages at full load 20 ms
CoolMod Power As per CoolMod table
Output Adjustment Range Manual: Multi-turn potentiometer. As per CoolMod table
Vtrim: As per CoolMod table
Overcurrent Protection Straight line with hiccup activation @ 35% Vo nom 110 130 150 %
Short Circuit Protection Yes, Autorecovery
OverTemperature Protection Yes, Autorecovery
Capacitive Load 10 mF
GALVANIC ISOLATION
Parameter Conditions/Description Min Nom Max Units
Input to Output Reinforced (2 x MOPP) 4000 VAC
Input to Case Basic (1 x MOPP) 1850 VAC
Output to Case Basic (1 x MOPP) 1850 VAC
Output to Output Basic (1 x MOPP) 1850 VAC
AUXILIARY OUTPUT
Parameter Conditions/Description Min Nom Max Units
Nominal Output Voltage Aux Voltage Option A
Aux Voltage Option B
11.76
4.75
12
5
12.24
5.25
V
V
Load Regulation ±2 %
Line Regulation For ±10% change from nominal line ±0.5 %
Maximum Output Current Aux Voltage Option A
Aux Voltage Option B
1.96
4.7
A
A
Maximum Output Capacitive
Load
1000 uF
Output Overcurrent Protection Hiccup 110 140 %
Short Circuit Protection Yes, Autorecovery
www.excelsys.com
3
RELIABILITY
Parameter Conditions/Description Min Nom Max Units
Reliability and MTBF MTBF of >400 kHours, Telecordia SR-332, Issue 1 CoolMod
CoolPac
0.52
1.08
Fpmh
Fpmh
Warranty 5 Years
CoolX600 NFF Series
ENVIRONMENTAL
Parameter Conditions/Description Min Nom Max Units
Operating Temperature Operates to specication below -20°C after 10 min warm-up -40 +70 °C
Storage Temperature -40 +85 °C
Derating See Derating Curves including note 2
Relative Humidity Non-condensing 5 95 %RH
Shock & Vibration MIL-STD810G Method 514.6
Altitude 5000 m
LEAKAGE CURRENTS
Parameter Conditions/Description
AC Leakage Current Input to earth ground Units
Normal condition (high line) Mains Voltage 264VAC/60Hz 44 µA
Single fault condition (high line) Mains Voltage 264VAC/60Hz 231 µA
Touch Current
Normal Condition Mains Voltage 264VAC/60Hz 83 µA
Single Fault Condition Mains Voltage 264VAC/60Hz 87 µA
EMC
Parameter Conditions/Description Criteria
Radiated Emissions EN 55011, EN 55022 and FCC, Class B
Conducted emissions EN 55011, EN 55022 and FCC, Class B
Power line harmonics EN 61000-3-2, Class A
Voltage icker EN 61000-3-3
ESD EN 61000-4-2, level 4, 8 kV contact, 15 kV air A
Radiated immunity EN 61000-4-3, level 2, 3 V/m A
Electrical fast transient EN 61000-4-4, level 4, ±4 kV B
Surge immunity EN 61000-4-5, level 4, 2 kV DM, 4 kV CM B
Conducted RF immunity EN 61000-4-6, level 2, 3 Vrms A
Power frequency magnetic eld EN 61000-4-8, level 2, 3 A/m A
STANDARDS & DIRECTIVES 1)
Parameter Conditions/Description
Safety Agency Approvals EN60601-1 3rd Edition, UL60601-1, CSA601,
EN60950 2nd Edition, CSA C22.2 No. 60950-1,
IEC/EN 60950-1, Edition 2 and
all national deviations UL 60950-1/CSA 22.2 No 60950-1, Edition 2;
5000 m (16,400 ft) altitude, 100 V to 240 ±10% Vac
IEC/EN 60601-1, Edition 3 and
all national deviations IEC 60601-1(2005), EN60601-1(2006)
ANSI/AAMI ES 60601-1(2005)
CAN/CSA C22.2 No. 60601-1 (2008);
5,000 m (16,400 ft) altitude, 100 V to 240 Vac ±10%
IEC 62368 Edition 2 IEC 62368-1 (2014) Edition 2
5000 m (16,400 ft) altitude,100 V to 240 ±10% Vac
IEC 60601-1-2 Edition 4 IEC 60601-1-2 (2014)
Protection class Class I
WEEE Waste Electrical and Electronic Equipment Directive (WEEE) 2002/96/EC
ROHS EU DIRECTIVE 2011/65/EC RoHS compliancy
REACH Compliant
1) Designed to support Type B and Type BF Applied Part End Product Requirements
www.excelsys.com
4
CoolX600 NFF Series
MECHANICAL DATA
Parameter Description
Dimensions (L x W x H) L x W x H 215.9mm x 114.3mm x 39.1mm 4.5” x 8.5” x 1U
Weight Nominal Weight: CoolPac + 4 x CoolMods 1Kg
Connectors Description Mating Connectors (if applicable)
AC/DC input terminal block TE 2-1437667-S, DINKLE DT-35-B07W-03
AC/DC IEC input (Option) IEC 320 Inlet
Main DC output terminal block M4 Screws
System Signal Connector J1005 Molex 87833-0831 8-way Locking Molex 51110-0860; Non Locking Molex 51110-
0850;
Crimp Terminal: Molex p/n 50394 or Molex 51110-0856
which includes Locking Tab & Polarization Keying
Output Signal Connectors J1001-1004 Molex 87833-0631 6-way Locking Molex 51110-0660; Non Locking Molex
51110-0650;
Crimp Terminal: Molex p/n 50394 or Molex 51110-0656
which includes Locking Tab & Polarization Keying
Output Sense Connectors J3 JST-S2BPH-K(LF)(SN) JST PHR2. Crimp Terminal JST BPH-002T-P.0.5S or
SPH-002T-P.05S
Auxiliary Output Connector J1 Molex 1041880210 2pin
CoolX Mechanical Drawings
215.90
3.00
39.10
6.00
162.00
19.55
122.00 31.50
36.15 42.00
24.25 24.25 24.25
114.30
CoolX can be mounted on its base, vertically or on its side. CoolX can also be mounted on the Excelsys DIN Rail accessory (Z744).
www.excelsys.com
12231.50
36.15 42.00
39.10
19.55
3.00
215.90
6.00
162.00
114.30
24.2524.25 24.25
5
CoolX600 NFF Series
CoolX 600 Eciency, Derang and Power Limit Curves
CoolX Derang
Conguring Your CoolX
www.excelsys.com
2) Enhanced thermal performance with system fans & base plate cooling. Consult Excelsys for details.
115V
230V
6
CoolX600 NFF Series
CoolX Connectors
CoolMod Connectors
CoolPac Connectors
www.excelsys.com
Specications are subject to change
REVJUNE16
Input Connectors
Standard (Screw Terminal) Opon 1 (IEC Terminal)